Problems that should be paid attention to during the overall calibration of the muffle furnace system
Problems that should be paid attention to during the overall calibration of the muffle furnace system
(1) The installation and wiring of the standard thermocouple furnace هغه وخت ترسره کیږي کله چې د مقاومت فرنس بند شي. وروسته له دې چې تارونه چک شوي، د تودوخې لوړولو لپاره د مقاومت فرنس فعال کیدی شي.
(2) The measuring end of the standard thermocouple should be placed near the sample analysis position, but the sample or the inner wall of the furnace should not be touched to prevent contamination or damage to the standard thermocouple.
(3) وروسته له دې چې د حرارت درجه د تودوخې کنټرول نقطې ته ورسیږي، د دې لپاره چې د فرنس د تودوخې یونیفورم او باثباته شي، د لوستلو دمخه د 30 دقیقو لپاره د تودوخې کنټرول اړین دی. څرنګه چې د مقاومت فرنس ثانوي میټر د ملی وولټ میټر نندارې سره همغږي نه دی، نو د ملیولټ میټر لوړ او ټیټ ارزښتونه باید د لوستلو هر پړاو لپاره ولوستل شي، او د ثانوي میټر لخوا ښودل شوي لوړ او ټیټ ارزښتونه باید مطابق ولوستل شي.
(4) د فرنس د تودوخې یونیفارم اندازه باید هغه وخت ترسره شي کله چې دا د تودوخې مستحکم حالت ته د نه بار کولو شرایطو لاندې راشي. د مدافعینو سره د مقاومت فرنسونو لپاره ، پرستار باید د اندازه کولو پرمهال په نورمال ډول کار وکړي.
(5) If the indication error is too large, check whether the furnace is working properly. For example, in order to prevent the stainless steel protection tube of the thermocouple of the control instrument from being shorted to the furnace wire, an insulating porcelain tube is placed on the thermocouple of the control instrument. In this way, the insulating porcelain tube should be cut short to avoid the actual length of the thermocouple of the control instrument being exposed. It is shorter than 8-10 times of the thermocouple protection tube; the hole of the control instrument thermocouple on the furnace wall should be sealed tightly to prevent heat loss in the nearby area. In the actual situation, certain measures can be taken to reduce the furnace temperature field correction value with a larger absolute value of the indication error.
(6) After calibration, in order to avoid damaging the thermocouple or shortening the life of the thermocouple, the standard thermocouple cannot be taken out quickly, and the furnace door cannot be opened quickly because the rapid heat and cold will damage the furnace.