site logo

The temperature of the muffle furnace shows the cause of instability

The temperature of the muffle furnace shows the cause of instability

1. The thermocouple electrode is in poor contact;

2. The thermocouple has an intermittent short circuit or open circuit;

3. The thermocouple is grounded intermittently;

4. The position of the thermocouple is constantly changing;

5. The compensation wire has grounding, intermittent short circuit or open circuit;

6. The temperature display instrument is faulty;

мера:

1. Reconnect the thermocouple;

2. Check the thermocouple to eliminate the fault point or replace it;

3. Eliminate grounding points;

4. Fix the thermocouple;

5. Check the compensation wire and eliminate the fault point;

6. Repair or replace the temperature display meter.