- 15
- Nov
How to correct the temperature of the muffle furnace?
How to correct the temperature of the sebopi sa muffle?
1. Set the temperature of the temperature control instrument to stabilize the temperature of the muffle furnace at this temperature;
2. Use a calibrated millivoltmeter to detect the output potential of the thermocouple;
3. According to the type of thermocouple, refer to the thermocouple temperature-potential comparison table of the model, such as (K-type nickel-chromium-nickel silicon) to find out the temperature value;
4. Measure the ambient temperature and perform cold junction compensation, that is, the table temperature plus the ambient temperature, which is approximately the actual temperature of the furnace.